Onicon F-2600 Manuale Utente Pagina 106

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 131
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 105
11451 Belcher Road South, Largo, FL 33773 • USA • Tel +1 (727) 447-6140 • Fax (727) 442-5699 • [email protected]
F-2600 & F-2700 Vortex Flow Meter Manual 02/15 - 0808-7 / 19204 Page 106
4. The vortex coefcient Ck may be incorrectly set. The Ck is a value in the equation used
to determine if a frequency represents a valid vortex signal given the uid density and
signal amplitude. In practice, the Ck value controls the adaptive lter, , setting. During
ow, view the f and  values in the rst level of the hidden diagnostics. The  value
should be approximately 10-20% higher than the f value. If you raise the Ck setting in
the Calibration Menu, then the  value will increase. The  is a low pass lter, so by
increasing it or lowering it, you can alter the range of frequencies that the meter will
accept. If the vortex signal is strong, the  value will increase to a large number – this is
correct.
Figure 54. Electronics Stack Sensor Connections
Figure 55. Remote Feed Through Board Sensor Connections
VORTEX
TEMPERATURE
PRESSURE
VORTEX
TEMPERATUREPRESSURE
Vedere la pagina 105
1 2 ... 101 102 103 104 105 106 107 108 109 110 111 ... 130 131

Commenti su questo manuale

Nessun commento