ONICON Inline Thermal Mass Flow Meters are provided with either threaded of anged process
connections. Threaded meters are provided with male NPT threads and anged meters are
provided with ANSI class 150 or class 300 anges. Flow meters should be installed in accordance
with the requirements of section 3.1 INSTALLATION SITE SELECTION.
ONICON does not supply gaskets for anged meters.
3.2.1 Threaded Meter
D
L
H
FLOW DIRECTION
Built-in flow conditioner
Output signal(s)
to control system
Optional
ONICON
Display
FLOW
FLOW DIRECTION
IMPORTANT NOTICE
Always use the maximum available straight run. When more than the minimum required straight
run is available place the meter such that the excess straight run is upstream of the meter location.
The ow meter must also be properly installed with respect to the direction of ow. Installing the
meter with the ow arrow pointing in the wrong direction will result in signicant errors in the
ow measurement.